Introduction
softWoRx
version
7.2.1 is a minor release providing increased reliability.
Changes and
Additions
Software
Defects Fixed:
- Export
to IN Carta: The Export to IN Carta tool is now
compatible with Acquire Ultra 1.2.0
- Task
Builder: The Volume Rendering task now properly
applies the chosen rendering method to all channels of the
image
- Resolve3D:
Experiments containing Reference Images and Panels are
now supported
- Image
Window: The unsupported Print option has been
removed from the File menu
Known
Problems and Workarounds
- Task
Builder -> OMX Align Image: When using the Legacy
Shift and Rotate method with Apply Z Alignment selected,
the Z alignment offsets are applied twice.
- Workaround:
Deselect Apply Z Alignment when using this method
or use the graphical softWoRx tool Process->Legacy->OMX
Image Registration and specify the Shift and
Rotate Registration Method which correctly
applies the Z alignment.
- DeltaVision
Panel File Format: Images collected using the Panel or
Plate Scan (with multiple panel sites) capabilities in
Resolve3D or AcquireUltra are not compatible with softWoRx
versions older than 7.0.0. If these files are stitched
on a system running 7.0.0 or later, they may then be
used with older software.
- DeltaVision
Acquisition: The Fast Acquisition
state can interfere with running a time-lapse Plate
Scan experiment. If Fast Acquisition is
toggled On and the Conventional Time Lapse
Fast Acquisition Option is toggled Off, only one
time point will be collected.
- Workaround:
Toggle Plate Scan Off temporarily (which
enables the Fast Acquisition check box),
switch Fast Acquisition Off, then switch Plate
Scan back On.
- Image Buffer Overwrite: When using the sCMOS
camera during a fast acquisition experiment with the Entire Scan
shutter mode enabled (the fastest frame rate mode) and
using a combination of large image sizes (1k x 1k and
larger) and very short time lapse values (< 5 ms), it is possible that the
instrument controller may not be able to process the
acquired images from the memory buffer fast enough. When this
happens, an image on the buffer waiting for download is
overwritten with a new image. The new image
is then captured as the image for the previous timepoint.
- Workaround:
Reduce the frame rate by using longer time lapse
values, reducing the image size, or reducing the
number of timepoints.
- Image
Window (Copy/Paste Scale): If a
"copied" set of display scale values exceed the Minimum
and/or Maximum values in the image to which the values
are "pasted," the scale values will be truncated to fit
within the Minimum/Maximum range.
- Workaround:
Manually type the values into the Min/Max/Exp field of the Scale
tool belonging to the window that will contain the
pasted values.
- Image
Window (Save...): If one
or more of the channels is toggled Off
in the Save tool, incorrect scaling of the output image
may result.
- Workaround:
Use the Copy Region tool from the main softWoRx
menu bar and send the output to a file instead of a
window.
- softWoRx: It has been
reported that individual user accounts seem to lose
permission to run softWoRx.
- Workaround: A possible
workaround is to remove (or rename) the file "user.swrc" in the .softworx_<hostname>
subdirectory of the user account's home directory.
- softWoRx: Some older
softWoRx versions or 3rd-party analysis tools can
have difficulty with the latest softWoRx default
data type.
- Workaround: Open the
image with softWoRx and choose Save
from the File menu of the image window. Choose Details...,
and change the Data Type from No Change to 2-byte
signed integer. Then, save a new copy of the
file.
- Stitch: If you
choose a subset of Z sections (from the Details...
dialog), Stitch may fail.
- Workaround: Restart softWoRx and
try again.
- Save
As Movie: Saving
a time-lapse image as a Movie when the window is in Montage
By Time viewing mode
may cause Save As Movie to crash.
- Workaround: Try choosing
the maximum that the image window time slider will go
and specify that number for the End time in
the Time: Start/End/Increment field of the
Save As Movie tool.
Then, select Do It.
- Save
As Movie: Using
the Movie tool to play the window as a movie while Save
As Movie is capturing images
can cause the Image window to crash.
- Workaround: Pause all other
movie tools prior to pushing Do It for the
Save As Movie tool.
- Save
As Movie:
Attempts to save more than one movie at a time can cause
crashes of either the Save As
Movie tool or the Image Window that it is using.
- Workaround: Ensure
any previous Save As Movie
processes have completed before starting a new one.
- Align
Image: This
tool can be unreliable when it is used on a
single-channel image.
- Workaround: There is
currently no known workaround for this issue.
- Calibration
File:
Calibration files are created with the file name
extension .dv_cal and may not
be viewable in softWoRx by default.
- Workaround:
Choose File -> Open from the softWoRx main
menu and change the File Types filter from *.dv to *
- Save
As Movie:
Graphical elements (like Line Profile graphics) are not
accurately synchronized with the image data when movie
frames are captured.
- Workaround: There is
currently no known workaround for this issue.
- Deconvolution: Attempting
to run two or more simultaneous multi-processor Deconvolution jobs on a
multi-core workstation may occasionally cause severe
slowing of all processing on the workstation.
- Workaround: If this
happens, cancel all but one of the jobs and allow it
to complete before starting any further jobs.
Release
Notes for Previous Versions
Version 7.2.0
softWoRx
version
7.2.0 is a major release providing improved functionality
and increased reliability.
Changes and
Additions
Software
Environment:
- softWoRx
is now supported on CentOS 5-7. This release is intended
to support DeltaVision (including Ultra, Elite, and other
legacy systems), and DeltaVision OMX processing.
Desktop
Environment:
- "Start
Acquire Ultra" and "Stop Acquire
Ultra" are now accessible from icons in the
desktop's "Kickoff Application Launcher" menu
Enhancements:
- Task
Builder: An option to save results as 4-byte
floating-point has been added to the Deconvolution task
options. This will appear in both the softWoRx
Task Builder and the Post-Acquisition Processing tasks in
the Run Experiment page in Resolve3D.
Software
Defects Fixed:
- Align
Image: The Align Image user interface has been
reworked to fix several issues related to specifying input
images
- Stitch:
The system configuration is updated by the Acquire
Ultra installation package to update the stage
polarity parameters for DeltaVision Ultra systems
- Export
Ultra to IN Carta: A problem with setting
the output folder to be the same as the experiment parent
folder has been resolved
- OMX
Align Image: The tool is more tolerant of
small image sizes and better communicates errors when
problems arise
- Image
Window: An issue that could have caused Image
Windows to fail when re-using them have been resolved
- Resolve3D:
Post Processing Options dialog box now can be closed
without selecting any tasks
- Enhance
Contrast: The tool no longer crashes with non-square
images
Version 7.1.0
softWoRx
version
7.1.0 is a major release providing improved functionality
and increased reliability.
Changes and
Additions
Acquisition (AcquireUltra):
AcquireUltra is the fully re-designed acquisition
client created to support the DV Ultra microscope system.
Software Environment:
- softWoRx
is now supported on CentOS 7.
Processing
and Analysis:
- Added capability
to export AcquireUltra plate scan to IN Carta (marked
as Beta)
- Added Flat Field
Correction of AcquireUltra images
Software
Defects Fixed:
- Resolve3D:
Panel Experiment canceled during first time point now
removes incomplete image file
- Resolve3D:
Critical / Kohler setting is now noted in the image log file
- Resolve3D:
Now writes the proper "1.0" photosensor value in the log
file when Use Photosensor is turned Off
- Resolve3D:
Panel experiment with Z sectioning now starts at the correct
Z position
- Resolve3D:
Deconvolve During Experiment now properly checks the time
lapse conditions
- Stitch:
Reduction Factor now properly applies to both X and Y axes
- Rotate3D:
Crash has been fixed
- AlignImage:
No longer crashes when an Image Window number is dragged in
to the Input field
- Save:
Saving a raw OMX file from an Image Window no longer
corrupts the Image Window
- Movie: The
movie tool no longer crashes softWoRx when running multiple
movies quickly
- Task Builder:
OMX Reconstruction now guards against illegal Weiner filter
values
- Create
Widefield From SI: The tool no
longer crashes when the output file name is changed
- OMX Align
Image: No longer fails when doing Z alignment step
- OMX Align
Image: Z alignment is now done before the XY alignment
to improve performance
- OMX Align
Image: Z alignment is no longer attempted if the image
has less than 2 Z sections
- OMX Align
Image: No longer quietly fails with 2048x2048 image
files
- OMX Align
Image: Processing speed has been improved
- OMX Align
Image: No longer fails at certain pixel sizes
Version 7.0.0
softWoRx
version
7.0.0 is a major release providing improved functionality
and increased reliability.
Changes and
Additions
Software
Environment:
- All of the
softWoRx software components running on the
workstation are 64-bit applications.
- Shared memory
sizes may be set much larger (15Gb) due to 64-bit memory
addressing.
- The
DeltaVision file format has been expanded to include a
"Panel" dimension to better accommodate plate scanning and
multi-panel acquisition.
Acquisition (Resolve3D):
- Paneling is
available for experiments with Z Sectioning, Multiple
Channels, Time Lapse, and Point Visiting.
- Time Lapse
imaging honors a per-channel Frequency setting for regular
channels and Reference Image channels.
- A
Multi-Channel Reference Image capability has been added to
the Experiment Design.
- Stage View is
now moveable and resizable.
- Stage View
thumbnails now have a resolution factor that can be
specified in Settings.
- Stage View
thumbnails can be adjusted for brightness.
- Image Scan
Sequence may be specified for Fast or Conventional imaging
modes.
- Fields of
View in Plate Scan experiments now use the file format's
Panel dimension, so separate files for fields are no
longer required.
DeltaVision
Processing Capabilities:
- The
Deconvolution tool has been improved to manage much larger
images and provide more rapid convergence for enhanced
ratio deconvolution.
- Task Builder
Process Chains may now be saved to be opened later.
- Stitching has
been added to the Task Builder.
OMX
Processing Capabilities:
- A new robust
OMX Image Alignment method has been added. Earlier tools
are still available in a "Legacy" submenu.
- OMX Image
Alignment Calibration plots markers on the input grid
image to indicate where usable alignment points are found.
Software Defects
Fixed
- DeltaVision
Acquisition: Panel experiments may now be
created with one row or one column.
- DeltaVision
Acquisition: Deconvolve During Time
Lapse experiments can now work along with experiment
Post-Processing tasks.
- DeltaVision
Acquisition: The Point List now properly
updates Z coordinates after running an experiment with
UltimateFocus.
- DeltaVision
Acquisition: Stage View thumbnails no longer
disappear when the lower left hits the left and bottom
borders of the view.
- DeltaVision
Acquisition: Size requirements for a Plate or Point
Visit experiment are calculated correctly and updated
during the experiment.
- OMX Image
Registration: The aligned image file no longer gets
corrupted when Z offsets are set to the same non-zero
value.
- OMX Image
Registration: Intensity Mean values are now
correct for images with Z shifts specified.
- OMX Image
Registration: Shift and Rotate method now works
properly from the Task Builder.
- OMX Image
Registration: The "most-recently used" parameter
files ar now being properly remembered by the software.
- OMX
Generate Widefield: This tool no longer fails when
specifying a Z subset.
- Arbitrary
Profile: The Full-Width Half Max
calculation has been improved.
- Stitch:
The interactive Stitch tool now has status reporting and
Cancel functionality.
- Stitch:
Enabling Remove Bright Pixels no longer causes the tool to
crash.
- Convolution:
Crash when re-using the tool has been fixed.
- Edit
Polygon: Closing the input Image Window is now
properly handled.
Version 6.5.2
softWoRx
version 6.5.2 is a minor release providing increased
reliability and improved functionality.
Changes and
Additions
- OMX SI
Reconstruction - Task Builder: Last-used Base OTF
files are now tracked by acquisition type (3Beam, 2Beam,
2Beam TIRF).
Software Defects
Fixed
- OMX SI
Reconstruction - Task Builder: All OTF files are now
properly validated against the acquisition type (3Beam,
2Beam, 2Beam TIRF) of the image to be reconstructed.
- Quick
Projection, Copy Region, Blend Colors, Image Fusion,
Filter 2D, Equalize Time Points: These tools now
properly construct output file name suggestions when writing
output to an image window.
- DeltaVision
Instrument Controller: The Instrument controller now
automatically attempts to re-establish communication with
the keypad if it becomes disconnected.
- DeltaVision
Instrument Controller: Resolve3D should no longer stop
writing new data to file in a single z, single channel, fast
acquisition (using shutter for each entire scan), and
time-lapse experiment.
Version 6.5.1
softWoRx
version
6.5.1 is a minor release providing increased reliability.
Software Defects
Fixed
- OMX PSF to
OTF: 3Beam PSF files created with pre-6.5.0
software are now properly identified and supported.
- OMX SI
Reconstruction - Task Builder: The per-channel
Wiener Filter constants are now properly used in 3D /
3Beam reconstructions.
- OMX SI
Reconstruction - Task Builder: k0 values should be
properly constrained to the range of -PI to +PI radians.
- OMX SI
Reconstruction - Task Builder: The Channel Specific
OTF Groups tool should now be properly initialized when
first opened.
- OMX Image
Registration and PSF to OTF tools: Input files
containing spaces in the folder or file name are now
supported.
- OMX SI
Reconstruction tool: An error message appears if the
input Image Window has been closed before choosing the Do
It button.
- Localization:
The Persistence Histogram is now properly updated when
changing the input Localization data file.
Version 6.5.0
softWoRx
version
6.5.0 provides enhanced functionality and increased
reliability.
Changes and
Additions
- OMX
SI Reconstruction: Support has been added for the
processing of 2-Beam 3D, 2D, and 2D TIRF data from OMX. The
SI Reconstruction tool will automatically determine the
imaging mode that was used to acquire the original data.
- OMX
SI Reconstruction: The Drawer dependency for
Channel-Specific options has been replaced with Emission
channel wavelength dependency.
- OMX
SI Reconstruction: Channel-Specific OTF assignments
are now organized as user-named OTF Groups.
- OMX SI Reconstruction: The log file produced by the 3
Beam Reconstruction tool
now provides a summary of refined k0 and
line-spacing parameters found in the k0 search.
- OMX
PSF to OTF: The tool now detects whether it's working
with 3 Beam or 2 Beam
PSF data and then generates a corresponding OTF file.
- OMX
Image Registration Calibration: This tool now puts
resultant calibration parameters files in a more visible
place (HOME_DIRECTORY/softWoRxConfig/)
with descriptive names.
- OMX
SI Reconstruction and Image Registration: Copies of
OTF and calibration parameter files are now saved in a
subdirectory (ParameterFiles/)
of the folder where the output data is stored.
- Task Builder: The OMX components
of the Task Builder have been updated to
the new processing style of the SI Reconstruction and
Image Registration tools.
- softWoRx:
A Close All menu option has been added to
the Window menu of the main softWoRx application to
clean up the Desktop by closing all (unsaved) image windows
and tools that are attached to them.
- softWoRx:
Tooltips have been added to the main menu system
to describe what each menu item does.
- Localization:
A Persistence (track length) Histogram has been added to
aid in evaluating fluorophore
tracking results.
- Localization:
The Localization Results viewer now allows you to filter
the view by maximum localization precision.
Software Defects
Fixed
- Localization:
Reconstructions referenced to previously-reconstructed
images are now rendered with proper coordinate systems.
- Localization:
Many problems related to file name handling and drag
and drop performance have been fixed.
Version 6.1.3
softWoRx
version
6.1.3 is a minor release providing enhanced
functionality and increased reliability.
Changes
and Additions
- Added 2D, 3Beam SI
Reconstruction capabilities to OMX processing
- Now includes a
MigrateOMXSettings tool to save and restore OMX-specific
user preference settings
Software
Defects Fixed
- Generate
Widefield from SI Image: This now works properly when
selecting a subset of channels, time points, and X,Y region.
It is required to specify the entire input Z section range,
however.
- SI
Reconstruction: The Reconstruction tool no longer
fails if a number greater than 99 was entered in a k0 angle
field.
- SI
Reconstruction: Channel-specific Wiener filter values
are now being handled properly when choosing a subset of
channels to process.
Version 6.1.2
softWoRx
version
6.1.2 is a minor release providing enhanced
functionality and increased reliability.
- Updated the driver
for the PCO sCMOS camera.
Version 6.1.1
softWoRx
version
6.1.1 is a minor release providing enhanced
functionality and increased reliability.
softWoRx
- Localization:
Localization Analysis is now roughly 2-3 times faster.
- Localization:
Localized fluorophores can now be tracked over time and
reconstructions can be based on the total intensity of
these grouped events.
- Localization:
Fiducial markers can now be detected and used to correct
images for stage X/Y drift. This method is in addition to
the existing Image Correlation drift correction.
- Arbitrary
Profile: A simple peak Full-Width Half-Max
calculation has been added.
- Resolve3D:
More icons have been added to the main control panel to
provide more direct access to several capabilities.
Version 6.1.0
softWoRx
version
6.1.0 is a minor release providing enhanced
functionality and increased reliability.
softWoRx
- Localization:
A suite of tools has been added to analyze
Localization microscopy images acquired on DeltaVision and
OMX systems. Included are methods to localize
single-molecule fluorophores, correct for sample drift,
and reconstruct super-resolution images from the
localization data. These tools will be available if
Localization is licensed on the workstation.
DeltaVision
Acquisition
- Localization:
Single-molecule switching experiments involving multiple
activation laser events are now supported. Tools have been
created to help design and interact with these experiments
as they are running. This experiment mode will be
available if Localization is licensed on the acquisition
workstation.
Version 6.0.0
softWoRx
version
6.0.0
is a major release providing enhanced functionality,
improved performance, and improved reliability. This edition
of softWoRx 6.0.0 is designed to be used with
DeltaVision microscope systems and OMXTM
processing workstations.
Changes
and Additions
softWoRx
- Task
Builder � File Copy: Changed
so that a simple file copy can be the only task. This can
be used to enable DropBox
capability in the open-source Omero
data management system.
- Task
Builder � Chromatic Correction: A
Chromatic Correction task has been added.
- Point
Values and Measure Distance: These
tools now optionally display values with sub-pixel
resolution to provide better support of Localization
Imaging.
- Image
Viewer:
Functionality has been added to the Montage Display to
allow for Color or Grayscale Channel viewing style.
- Save As TIFF: Support
for 32-bit floating point format has been added as an
option.
- Trails
Movie: An
option has been added to do a simple Maximum Intensity
calculation over time when creating a trails movie.
DeltaVision
Acquisition
- Scientific
CMOS: Added support
for the scientific CMOS camera implemented for its
sensitivity, CCD size, and frame rates up to 400 fps for
specific experiment types.
- Microscope
Control: Added support
for the latest-generation Instrument Controller,
including its dramatically increased data transfer
speeds.
- Experiment
Designer: New
workflow layout for Design/Run Experiment tool.
- Shutter
Selection: All
illumination source switching is now handled
automatically.
- Panel
Collection Focus: Autofocus
and focus maintenance (UltimateFocus)
can be applied during panel collection.
- Post-Acquisition
Processing: Tasks may
now be as minimal as copying the image and log files to a
different location (enables use of OMERO "DropBox" capability, for
instance).
- Find
Exposure: The Find
Exposure tool is now better optimized for bright InsightSSITM
illumination.
OMX Support Changes
- Warp
Image Alignment: A new
algorithm has been implemented to provide more accurate,
non-uniform X/Y channel-to-channel image alignment.
Z-shifts (expressed in microns) are entered manually and
are not restricted to whole-pixel increments.
- Enhancements
for Time Lapse Data: These
include proper timestamp handling in the SI Reconstruction
and options for enabling/disabling k0 angle recalculation
for each time point.
- Create
Widefield Image: A new tool
has been added to create a widefield
image from 3D SI data.
- Migrate OMX
Settings: A tool
has been added for migrating OMX-related softWoRx settings
between workstations.
Version
5.5.0
softWoRx
version
5.5.0
is a major release providing enhanced functionality,
improved performance, and increased reliability.
Changes
and Additions
Usability,
Visualization, and Analysis
- File
Chooser: The Open
Image File Chooser now shows the file name list in a way
that can be sorted by Date or File Name, ascending or
descending.
- File
Chooser: You can
now select a text file (an R3D log
file for instance) with the Open Image File
Chooser and it will open the file in a text editor.
- Task
Builder: You can
now specify a directory (network storage device, for
instance) to which input images are to be moved before
processing them.
- Task
Builder: The OTF
file to be used for Deconvolution
can now be specified in the Deconvolution
Task Options.
- OMX
Support: OMX SI
Reconstruction, Image Alignment, and OMX OTF tools are now
available for appropriately-licensed systems.
Software
Defects Fixed
- Montage
Display: The
Image Window now properly reports coordinates when in
Montage Display mode.
- Camera
Gain: Due to
many complications and unexpected consequences, camera
gain can no longer be set to different values for
different channels.