To design a Multiplexed Wavelength experiment:
From the Resolve3D main menu, click the Experiment button to open the Design/Run Experiment window.
If the Multiplexed Wavelength option is enabled on your DeltaVision system, you’ll see the Multiplexed tab in the Design/Run Experiment window. Click on the Multiplexed tab to view the options for Multiplexed Wavelength experiments.
From the Multiplexed tab, select the Do Multiplexed Channel Imaging check box.
If the currently active filter set is not Multiplex capable, an error window is displayed. You will need to change the active filter set to continue. Press OK to return to the Design/Run Experiment window. To change the active filter set for Multiplexed Wavelength experiments, see the procedure for activating a multiplexed wavelength filter set in “Setting Up the Multiplexed Wavelength Option.”
If the currently active filter set is Multiplex capable, the Design/Run Experiment window is displayed.
At this point, you have completed the initial portion of the Multiplexed Wavelength experiment design setup. You should now continue with the standard steps for the remainder of the experiment design, such as Sectioning, Timelapse, and so on.
Note: As soon as you activate the Do Multiplexed Channel Imaging check box, all conventional imaging settings are disabled. This is also true of all multiplexed settings when you reactivate conventional imaging. |