Localization Phase Input Parameters
To determine the sub-pixel positions of the fluorophores
in a data set:
1. From the softWoRx main menu, select Measure
| Localization Image Analysis. The Localization Image Analysis
window is displayed.
2. Click the Localization
tab to activate the Localization portion of the window.
3. Click the Input
Image button to browse and select the image file to use for analysis.
Alternatively, you can click and drag an image window number into this
field. This file must be a DeltaVision image file (*.dv)
4. To use only a portion of the input image, click
the Select Region button and then
click and drag the mouse to specify the part of the image you want to
analyze. Alternatively, you can use the Details
button to define a subset to analyze.
5. Activate the Auto-increment
file names check box to automatically increment the _LOC.txt
file names for the experiment.
6. The Localization
Data field is automatically filled with your original localization
file name (_LOC.txt).
Click the Localization Data button
if you want to browse for and select a different localization file.
7. In the PSF size
factor field, enter the PSF size factor in units of pixels. The
PSF size factor is the predicted, or measured, value of the standard deviation
of the Gaussian fit to the point spread function of the microscope for
the experimental conditions. It may be necessary to measure the PSF size
factor directly using the PSF Width tool under the Measure menu. Use the
Set to default button to set this
field to the theoretical value for the chosen experiment conditions.
8. Enter the desired local maximum factor into the
Local maximum factor field. The
local maximum factor should be a value from 0.0001 to 0.5. The local maximum
factor is 1 minus the probability that a detected peak is significant
above the background distribution (also known as an alpha value). Therefore,
the local maximum factor will have typical values of 0.1, 0.05, and 0.01.
The default value is 0.05 and larger values lead to more detected peaks.
9. Click the Test
Localization Parameters button to test the selected parameters
on a small sample of the data. To use the Test
Localization Parameters button, the raw DeltaVision image file
must be open and used as the input to the Input
Image field (drag a window number into the Input
Image field as in Step 3 above).
10. By default, the input localization parameters will
be tested on frames 1-20 of the dataset. To change these defaults, click
on the Test Options button. The
test options window is displayed.
Use the Test Start
Frame field to select the frame where the test of localization
parameters will begin. Use the Number
of Test Frames field to enter the total number of frames in the
localization parameters test. If desired, deselect the Start
test at current frame check box to have the test start at the Test
Start frame. By default this box is activated meaning that the test will
start on the current frame of the raw data and proceed for the number
of test frames. Click the Done
button to close the Test Options window and return to the previous window.
11. If desired, activate the Show
Localization Viewer check box, which will display the Localization
Viewer once the data has been analyzed.
12. When you are satisfied with the selected detection
parameters, click the Run the Localization
button to process the indicated frames.
13. The selected detection parameters can be used to
analyze data sets in a queue. To run the localization as a background
task, click the Add Localization to the
Queue button.
To evaluate the selected parameters, you can compare the detected fluorophore
positions to the original acquired image using The
Localization Results Viewer.
Related Topics
Localization Phase
Correction Phase
Tracking Phase
Reconstruction Phase
The Localization
Results Viewer
The
Localization Precision Histogram