Use options on the Lasers tab of the Resolve3D Settings window to configure and align a laser for a photokinetics experiment.
Open the Laser settings options by clicking the Settings
icon on the Resolve3D window and clicking the Lasers
tab.
Field |
Description |
Module Status |
Indicates whether the optional attached laser module is Off or On. |
Laser |
View which lasers are currently installed in the Laser Module Chassis. |
Status |
View the states of the Lasers. Status displays Off, Standby, Warming Up, or Ready for each laser. You cannot use a laser in an experiment until Status displays Ready. |
Power (%) |
The control voltage value that determines laser power. A typical value is 20%.
The Power value is the percentage of maximum control voltage. Increasing control voltage increases laser power; however, control voltage is not linearly related to laser power. |
Laser to test |
Use the drop-down list in this field to select a specific laser on which to run PK tests. |
Show spot alignment target |
Activate this check box to include a target alignment reticle on the displayed test image as shown below. |
Image The Spot |
Selecting this button opens the laser shutter and acquires a single image of the laser spot. The event duration is set using the Exposure field in Resolve3D. |
Continuous |
Selecting this button opens the laser shutter and acquires continuous images of the laser spot. Select Stop Imaging to discontinue the imaging process. |
Do Test Bleach |
Perform a bleach test using the current Power and PK Duration settings. |
% Bleach |
The value entered in this field displays a before and after comparison of the intensities at the bleach spot: 100x (Signal After/Signal Before). |
Background |
The value in this field provides a measurement of the signal (intensity value) after the bleach event. |
TIRF Illumination Settings |
Selecting
this button opens the TIRF Illumination Settings window from which
the following actions are available: TIRF
Beam Depth – allows you to select a laser to adjust its
evanescent wave depth. Block
– sets the TIRF depth adjuster to a position that prohibits illumination
of the sample. Epifluorescence
– sets the TIRF depth adjuster for epifluorescence (the highest
value of the TIRF depth slider). Restore
TIRF Depth – restores the TIRF depth adjuster to the last-used
TIRF position. Back
Aperture Focus for Current Lens – focuses the laser on
the back aperture of the objective resulting in a collimated illumination
laser and an even TIRF depth across the field of view. This value
should be set once for each installed objective. To make sure
this position is set correctly for imaging, verify the proper
objective is selected from the Resolve3D drop-down menu. The Locked check box (activated
by default) keeps the slider from being inadvertently moved. Laser Path Splitter – sets the percent of laser light to be directed to the TIRF illumination path versus the PK path.
Use Beam Concentrator - enables the laser beam concentrator when check box is selected. |
|
|