Setting Up Localization
To set up and prepare for a localization experiment:
1. From the Resolve3D window, click the Experiment
icon. The Design/Run Experiment window is displayed.
2. Select the Design
tab to activate it.
3. Select the Localization
tab on the left side of the window.
4. In the Experiment
name field, you can either accept the default name, or rename the
experiment in order to retain the settings for future experiments.
5. Click on the Laser
Module Settings button. The Resolve3D Settings window is displayed
with the Lasers tab activated.
Use this window and the TIRF Illumination Settings window to set up the
DeltaVision system as you normally would to perform a TIRF-based experiment.
This procedure is described in the “Imaging Using TIRF” section in Chapter
8 of the DeltaVision Imaging System User’s Manual.
Note: In the TIRF Illumination
Settings window, move the Laser Path Splitter slider all the way
to the right (for 100% TIRF). |
6. In the Reporter Imaging Setup portion of the Design/Run
Experiment window, use the drop-down menu in the Exp
field to set the exposure time. Then enter your selected EX/EM pair using
the drop-down menus for the EX Laser
and the EM Filter.
Note: When you make a
selection from the EX Laser
drop-down list, the rest of the fields in the Reporter Imaging
Setup portion of the window are populated automatically with the
last values used. |
7. To leave the excitation laser turned on throughout
the experiment, activate the Continuous
Illumination check box. Leaving the laser on will help to keep
photoswitchable dyes in their “off” states during acquisition so that
only a few fluorophores are on in any given frame. Though the photophysics
are not completely understood, it may be useful to pulse the excitation
laser when using certain fluorophores, such as photo-convertible fluorescent
proteins, to help avoid photobleaching or unwanted transitions into triplet
states.
8. Activate the Switch
fluorophores Off check box to use the Reporter Imaging laser to
switch all fluorophores to an “off” state before the first activation.
With this check box activated, you can adjust the Laser
power, Pre-Activation Images,
and Event duration (secs) fields
as desired.
Note: Pre-activation images
are used to confirm that the majority of the fluorophores have
been switched to the “off” position. |
In some cases, it may be useful to initially switch the
fluorophores to an “off” state prior to data acquisition. This can be
very useful when dyes have been used for labeling. The user must be careful
not to subject the sample to an excessively long pulse duration, which
could photobleach the sample. It may be necessary to empirically determine
the optimal duration for each sample type.
9. In the Activation/Imaging Sequence portion of the
Design/Run Experiment window, use the drop-down list in the Activation
field to select each laser wavelength (up to three lasers) for the activation
laser(s). To the left of the Activation
field, click the + button to add
another laser or click the - button
to remove a laser.
10. In the Power
field, enter the percent of laser power to use for this wavelength. This
field should be set at a minimum of 10% for laser stability.
11. In the Duration
field, use the drop-down menu to select the duration for each laser pulse.
Note: If you have the
Continuous Illumination
check box activated, you cannot set the duration for any of the
activation lasers to Continuous. |
12. In the Sequence
repeats field, enter the number of times you want to repeat the
entire sequence.
13. Use the settings in the UltimateFocus Setup portion
of the window to set up UltimateFocus for the experiment. Enter your selection
for how often you want UltimateFocus to perform a Z-focus maintenance
correction in the UltimateFocus every
field. This is the number of sequences to wait before performing UltimateFocus.
14. In the Move threshold
field, enter your selection for the move threshold. Any drift greater
than the value set in this field is corrected.
15. In the Maximum
iterations field, enter the maximum number of iterations for UltimateFocus.
This selection represents the maximum number of times the UltimateFocus
function will be performed to find optimal focus during a single correction
event.
Related Topics
Using the Beam
Concentrator
Calibrating
the Camera for Localization
Using
the Localization Progress Monitor
PSF Width Tool
Localization Phases
The Localization
Results Viewer