Use options on the Resolve3D Settings Imaging tab to select camera settings before you acquire an image.
Open the Imaging settings options by clicking the Settings
icon on the Resolve3D window and clicking the Imaging
tab.
Use |
To... |
Camera |
Specify which camera to use. |
Frames to average |
Specify the number of successive camera images to average into a result that is displayed in the window. (Similar to the AVG macro command.) |
Gain
|
Choose a gain value from the list of those supported for the currently selected camera. |
Transfer speed |
Choose a transfer speed from the list of those supported for the currently selected camera. (Higher speeds may boost performance at the expense of image noise.) |
Target temperature |
Read the cooled CCD camera's target temperature. This can be changed only by modifying the appropriate configuration file on the Instrument Controller. |
Current temperature |
Read the current temperature. Resolve3D monitors the temperature and updates this value occasionally. To force a request for the current temperature, click Refresh. |
Use photosensor |
Enable or disable use of the photosensor. |
|
|